EBI 12-T421 Validation logger to measure in steam sterilizer
Product number:
1340-6630
Measurement up to +400 °C
This data logger is often used for thermal measurement during validation. Its two 500 mm long bendable metal probes offer maximum flexibility in the placement of the temperature probes.
Due to its extended measurement range, the logger can also be used for quality control during roasting processes. To ensure that the logger is not damaged by the high temperatures, it must be protected from the effects of temperatures above +150 °C. The logger can be used in a variety of applications. For this purpose, you will find the TIB2 temperature protection box in the accessories.
Benefits:
- large measuring range
- bendable metal probe
- high accuracy
- Realtime data transmission
- wireless measurement
Applications:
- Validation logger for temperature measurement in the steam sterilizer, WD and endoscope washer
- Validation in the hot air sterilizer
- Validation of baking processes
Datasheet
- Data memory
- 2 x 50,160 measurement values
- Sensor
- Pt 1,000, class A
- Weight
- approx. 110 g
- Accuracy
- ± 1,5 °C (-200 °C … -85 °C) ± 0,5 °C (-85 °C … -40 °C) ± 0,2 °C (-40 °C … 0 °C) ± 0,1 °C (0 °C ... +120 °C) ± 0,05 °C (+120 °C … +140 °C) ± 0,1 °C (+140 °C … +150 °C) ± 0,5 °C (+150 °C … +250 °C) ± 0,8 °C (+250 °C … +400 °C)
- Protection class
- IP68
- Dimensions
- (Ø x H) 48 mm x 24 mm - Needle length 500 mm
- Resolution
- 0,01 °C
- Certificate
- Factory Calibration: -20 °C, 0 °C, +60 °C, +134 °C and +250 °C
- Measured variables
- Temperature
- Intervall
- 250 msec ... 24 h
- Measuring range
- -200 °C ... +400 °C
- Storage temperature
- -20 °C ... +70 °C
-
• Developed for real time monitoring
• AL 112 is a cable with 3 m length
• Antenna placement in the field of the door seal of a sterilizer or washer disinfector
• Antenna steam-tight
• It is possible by a sealing to introduce it in the chamberProduct number: 1340-6007 -
• Usable from +150 °C ... +400 °C
• Thermal protection of data loggers
• Stainless steelProduct number: 1340-1892 -
For securing the EBI 12 data loggers in process you need the AL 121 opening tool.Product number: 1248-0122
-
4-port Interface IF 200
Software Winlog.validation
USB connection
Colored LEDs signaling programming, readout and incorrect development
Includes antenna
Product number: 1340-6068 -
4-port Interface IF 200
Software Winlog.pro
USB connection
Colored LEDs signaling programming,readout and incorrect development
Includes antenna
Product number: 1340-6062 -
2-port Interface IF 100 Software
Winlog.pro
USB connection
Colored LEDs signaling programming, readout and incorrect development
Product number: 1340-6061 -
USB-connector
Colored LED signals programming,
readout and faulty execution
Including Antenna AL 111
Software Winlog.medProduct number: 1340-6095 -
1-port Interface IF 150
Software Winlog.med
USB connection Colored
LEDs signaling programming, readout and incorrect development
Product number: 1340-6193
-
For EBI 12 + EBI 16 Consisting of 3 batteries, 6 O-rings with grease.Product number: 1100-0118
-
Consisting of 3 batteries, 6 O-rings with grease and opening tool EBI 12Product number: 1100-0130
-
For transport of systainers Fixed seat of the systainers on the trolleyProduct number: 1248-0127
-
- Combined
- Modular construction
- Loggers are safely stored
Product number: 1248-0128 -
Especially for validation sets Insert for EBI 12 and InterfaceProduct number: 1248-1100
-
ISO Calibration for temperature- validation- and process logger at 5 standard points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Product number: 1030-2205
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at three points. Calibration service for data logger of the following product series:EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3203
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at four points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3204
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at five points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3205
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at six points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3206
- 2014/34/EU ATEX Directive
- ISO 17665 Sterilization of health care products — Moist heat
- ISO 15883 Washer Disinfectors
- 2014/53/EU Radio Equipment Directive - RED
- (EC) 1935/2004 Materials and articles which come into contact with foodstuffs