EBI 12-TP222 Validation logger to measure in autoclaves

Product number: 1340-6653

Easy to place the bendable metal probe for temperature measurement

Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its bendable metal probes. The very thin temperature sensors with a diameter of 1.5 mm at the sensor tip enable a very fast response time. With this logger and the AL 126 Endoscope Flow Control Adapter, regular routine checks in RDG-E are easy to perform.

Benefits:

  • bendable metal probe
  • high accuracy
  • Realtime data transmission
  • fast response time
  • sterilizable data logger

Applications:
  • Validation logger for Measurement in the steam sterilizer, WD and endoscope washer

Datasheet

NEW: Information package "Professional data loggers"

You still need a more comprehensive overview of the possibilities and requirements of the data logger, references to the service, certificates or the suitable software? Then read the comprehensive information package for the food sector or the medical/pharmaceutical sector here.
Go to information package
Data memory
2 x 33.440 valores
Weight
aprox. 120 g
Accuracy
±0,1 °C (0 °C ... +120 °C) ±0,05 °C (+120 °C ... +140 °C) ±0,1 °C (+140 °C ... +150 °C)
±15 mbar (1 mbar ... 50 mbar), ±10 mbar (50 mbar … 150 mbar), ±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar), ± 15 mbar (3.250 mbar ... 3.500 mbar),
± 0,5 % FS (3.500 mbar ... 4.000 mbar)
Protection class
IP68
Dimensions
ø 48 mm x 32 mm
Resolution
0,01 °C 1 mbar
Certificate
Certificado de calibración de fábrica
Measured variables
Temperatura, presión
Intervall
250 ms ... 24 h
Battery
3.6 V lithium battery exchangeable
Measuring range
Temperatura: 0 °C … +150 °C Presión: 1 … 4.000 mbar
Storage temperature
-20 °C ... +70 °C
    • Developed for real time monitoring
    • AL 112 is a cable with 3 m length
    • Antenna placement in the field of the door seal of a sterilizer or washer disinfector
    • Antenna steam-tight
    • It is possible by a sealing to introduce it in the chamber
    Product number: 1340-6007
  • The test adapter offers a simple solution in connection with cleaning indicator (test soil) to perform a routine test in relation to cleaning performance as well as performance qualification during validation processes according to national standards. In addition to AL 126 we recommend also to use a combination of pressure and temperature logger.
    The set contains: 
    • Test adapter 
    • Luer-Lock connection to connect a
      pressure logger 
    • 2 x Tube adapter 4 mm 
    • 1 x clamp connection 1.5 mm 
    • 2 x O-rings for fitting test soil

    Product number: 1248-0126
  • for opening the data logger, e.g. for battery change. The set consists of 2 parts

    Product number: 1100-0131
  • Eyelet ring incl. sealing rings for EBI 12 to mount the data logger
    Product number: 1248-0122
  • AL 285 Logger Check In combination with the software Winlog.validation version >= 3.62 the AL 285 offers you a simple, on site functional check of the measuring channels.
    Product number: 1248-0285
  • 4-port Interface IF 200
    Software Winlog.validation
    USB connection
    Colored LEDs signaling programming, readout and incorrect development
    Includes antenna

    Product number: 1340-6068
  • 2-port Interface IF 100
    Software Winlog.med
    USB connection
    Colored LEDs signaling programming, readout and incorrect development

    Product number: 1340-6064
  • 4-port Interface IF 200
    Software Winlog.pro
    USB connection
    Colored LEDs signaling programming,readout and incorrect development
    Includes antenna

    Product number: 1340-6062
  • USB-connector
    Colored LED signals programming,
    readout and faulty execution
    Including Antenna AL 111
    Software Winlog.med

    Product number: 1340-6095
  • 1-port Interface IF 150
    Software Winlog.med
    USB connection Colored
    LEDs signaling programming, readout and incorrect development

    Product number: 1340-6193
  • For transport of systainers Fixed seat of the systainers on the trolley
    Product number: 1248-0127
  • Especially for validation sets Insert for EBI 12 and Interface
    Product number: 1248-1100
    • Combined
    • Modular construction
    • Loggers are safely stored
    Product number: 1248-0128
  • ISO Calibration for pressure- temperature data logger, humidity- and conductivity data logger at 3 standard temperature points and the calibration of the additional measurand of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.

    Product number: 1030-2223
  • Accredited calibration according to DAkkS guidelines for pressure- temperature data logger, humidity- and conductivity data logger at three temperature points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.
    Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.
    Product number: 1030-3223
  • Accredited calibration according to DAkkS guidelines for pressure- temperature data logger, humidity- and conductivity data logger at four temperature points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.
    Consisting of an accedited calibration according to DIN EN ISO / IEC 17025.
    Product number: 1030-3224
  • Accredited calibration according to DAkkS guidelines for pressure- temperature data logger, humidity- and conductivity data logger at five temperature points. 
    Calibration service for pressure- temperature data logger of the following product series: 
    EBI 12, EBI 11, EBI 10, EBI 100
    Consisting of an accedited calibration according to DIN EN ISO / IEC 17025. 
    Product number: 1030-3225
  • Accredited calibration according to DAkkS guidelines for pressure- temperature data logger, humidity- and conductivity data logger at six temperature points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.
    Consisting of an accedited calibration according to DIN EN ISO / IEC 17025.
    Product number: 1030-3226
  • ISO 17665 Sterilization of health care products — Moist heat
  • ISO 15883 Washer Disinfectors
  • 2014/53/EU Radio Equipment Directive - RED
  • (CE) N° 1935/2004 Materiales y objetos que entran en contacto con los alimentos

NEW: Information package "Professional data loggers"

Please fill out the form below and the download will start instantly.
Zertifikate