EBI 12-TP290 Precision Pressure Logger
high precision pressure measurement down to 0.1 mbar
Product information
In the VH2O2 sterilizer, the generation of a fine vacuum is a prerequisite for the success of the sterilization process. For a routine check but also for measurement during validation, this data logger with its high accuracy is available.
The data logger must never be used in a steam sterilizer or otherwise exposed to an absolute pressure >2,000 mbar.
Benefits:
- high accuracy
- robust data logger
- fast response time
- Realtime data transmission
- Sterilization mit VH2O2
Datasheet
Technical Data
- Data memory
- 33,440 measurement values
- Sensor
- Pt 1000, piezoresistive pressure sensor
- Weight
- Aprox. 120 g
- Accuracy
- Pressure: 0.25 mbar (0.1 to 50 mbar) ±5 % of measured value (50 to 100mbar) 1 % FS (100 to 1050 mbar) Temperature: ±0.1 °C (0 °C to +85 °C)
- Protection class
- IP68
- Dimensions
- ø 48 mm x 35 mm - Needle length 40 mm
- Resolution
- Pressure: 0,1 mbar Temperature: 0,01°C
- Certificate
- Factory Calibration: 0 °C and +60 °C
- Measured variables
- Pressure, Temperature
- Intervall
- 250 ms...24 h.
- Battery
- Lithium cell, 3.6 V, user replaceable
- Measuring range
- Pressure: 0,1 mbar ... 1050 mbar.Temperature: 0°C ... +85°C