EBI 12-TP460 Registrador de validación
bendable sensor for placement in cavities
EBI 12-TP460
Registrador de validación
Número de producto:
1340-6648
WEEE-Reg.-Nr. DE 29884659
Información sobre el producto
Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning.
With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers.
The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time.
When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.
Benefits:
- highly flexible cable probes
- high accuracy
- Realtime data transmission
- fast response time
- sterilizable data logger
Applications:
- Validation logger for Measurement in the steam sterilizer
NEW: Information package "Professional data loggers"
You still need a more comprehensive overview of the possibilities and requirements of the data logger, references to the service, certificates or the suitable software? Then read the comprehensive information package for the food sector or the medical/pharmaceutical sector here.Datos technicos
- Capacidad de almacenamiento
- 4 x 20,064 valores
- Pesa
- aprox. 120 g
- Precisión
- ±0,1 °C (0 °C ... +120 °C) ±0,05 °C (+120 °C ... +140 °C) ±0,1 °C (+140 °C ... +150 °C)
±15 mbar (1 mbar ... 50 mbar), ±10 mbar (50 mbar … 150 mbar), ±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar), ± 15 mbar (3.250 mbar ... 3.500 mbar),
± 0,5 % FS (3.500 mbar ... 4.000 mbar) - Clase de protección
- IP68
- Dimensiones
- ø 48 mm x 24 mm
- Resolución
- 0.01 °C 1 mbar
- Certificado
- Certificado de calibración de fábrica
- Variables medidas
- Temperatura, presión
- Intervalo de medida
- 250 ms ... 24 h
- Batería
- 3.6 V lithium battery exchangeable
- Campo de medición
- Temperatura 0 °C … +150 °C Presión 1 … 4.000 mbar
- Temperatura de almacenamiento
- -20 °C ... +70 °C
Accessories
Software & Interfaces
Cambio de batería y & Mantenimiento
Maletín de transporte
Calibration
NEW: Information package "Professional data loggers"
Please fill out the form below and the download will start instantly.
Standards and Directives
• (EC) 1935/2004 Materials and articles which come into contact with foodstuffs• ISO 17665 Sterilization of health care products — Moist heat
• 2014/53/EU Radio Equipment Directive - RED