SL 1521 Supervisión de esterilizadores de vapor

Número de producto: 1250-1521

for steam sterilizers certified according to ISO 11140-4

EBI 16-Set SL 1521 contiene:

  • 1 x registrador de datos Bowie&Dick EBI 16
  • 1 x Interfaz EBI IF 150
  • 1 x Software Winlog.med
  • Instrucciones de manejo y maletin

NEW: Information package "Professional data loggers"

You still need a more comprehensive overview of the possibilities and requirements of the data logger, references to the service, certificates or the suitable software? Then read the comprehensive information package for the food sector or the medical/pharmaceutical sector here.
Go to information package
  • L'enregistreur de données EBI 16 est un système de mesure destiné au contrôle des autoclaves.
    Outre la surveillance des paramètres de stérilisation pertinents que sont le vide, la durée d'équilibrage, le temps de maintien et la température de stérilisation, ce système permet de vérifier l'entrée de vapeur de la même manière que dans le cadre d'un test Bowie & Dick. La température à l'intérieur d'une chambre de mesure est en outre comparée avec la température extérieure.Un écart de température signifie que l'entrée de vapeur est insuffisante. En fonction de l'importance de la différence et de l'allure dans le temps des courbes de températures, un message indiquant la raison de l'insuffisance de l'entrée de vapeur (fuite, vide insuffisant, injection d'air) peut être affichée. Par principe, la même procédure est utilisée ici que lors de la détermination de l'entrée de vapeur effectuée à l'aide d'un PCD (process challenge device) à hélice. Toutefois, une mesure physique de la température est utilisée ici, et non pas un indicateur chimique. L'interprétation des différences de température s'effectue automatiquement par lecture des valeurs mesurées par le logiciel (Winlog.med, Winlog.pro) et le résultat est affiché, le cas échéant avec l'origine du défaut.
    Número de producto: 1340-6697
  • Interfaz de 1 pliegue IF 150
    Software Winlog.med
    Conexión USB
    El LED de color señala la programación, la lectura y la ejecución defectuosa

    Número de producto: 1340-6193
  • Para transportar el Systainer.
    Número de producto: 1248-0127
  • Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

    The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

    The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

    •  For routine checks and validation
    • High-performance report and evaluation software for the stringent requirements of validation and qualification in the pharmaceutical and medical sectors
    • User-friendly
    • Highly accurate measurements
    • Automatic report generation
    • Automatic user-defined calculations
    • Automatic recognition of process cycles
    • Creation of user-defined templates for specific devices and thermal processes
    • Three-dimensional display of sensor placement
    • FDA 21 CFR Part 11 functionality also possible
    • IQ- / OQ documentation available
    • Automatic completion of validation processes
    • Current version: 4.0

    ebro® offers exactly the right solution and the evaluation software you need for your application.

    Currently, the products of the device series EBI 1, EBI 2, EBI 15, EBI 30 are still supported by the software. However, the software is no longer tested with the units. 

    We will be happy to make you an offer – please call us:

    +49 841 95478-22, +49 841 95478-23, +49 841 95478-37


    Número de producto: 1340-2470
    • Supports: EBI 10, EBI 12, EBI 11, EBI 16, EBI 20, EBI 40, EBI 100, EBI 300, EBI 310
    • For routine checks
    • User friendly High-precision measurements
    • Automatic report generation
    • Automatic user-defined calculations
    • Automatic recognition of process cycles
    • Creation of user-defined templates for specific devices and thermal processes
    • Three-dimensional display of sensor placement
    • FDA 21 CFR Part 11 functionality also possible
    • Current version: 3.71

    ebro® offers exactly the right solution and the evaluation software you need for your application.