EBI 11-T236 Temperature validation logger
Product number:
1340-6271
Validation logger for temperature measurement in the doctors surgeries
The EBI 11-T236 validation logger was specially developed for the validation of reprocessing of medical devices in the tertiary sector. It is adapted to the validation concept for the Sirona DAC Universal and Melag Careclave units.
Benefits:
Applications:
- Data memory
- 15.000 measurement value
- Weight
- ca. 45 g
- Accuracy
- ±0,1 °C
- Protection class
- IP 68
- Dimensions
- ø 16,5 x 24 mm (without probe) - Needle length 86 mm
- Resolution
- 0,01 °C
- Certificate
- Factory Calibration: 0 °C, +60 °C and +134 °C
- Measured variables
- Temperature
- Intervall
- 1s...24h
- Battery
- Battery 2 x 3 V
- Measuring range
- -30 °C...+150 °C
- Storage temperature
- -20 °C...+70 °C
-
With the Adapter set is an independent process qualification of a DAC Universal MK III and MK IV incombination with ebro EBI 11 data loggers possible. No modification on the customers lid necessary.The set contains:• 1 x Logger mount
• 1 x PCD test body incl. adapter for ISO, Sirona and KaVo
• Carrying case „SYSTAINER“Product number: 1248-3305 -
AL 285 Logger Check In combination with the software Winlog.validation version >= 3.62 the AL 285 offers you a simple, on site functional check of the measuring channels.Product number: 1248-0285
-
4-port Interface IF 300
Software Winlog.validation
USB connection
Colored LED signalling programming, readout and incorrect development
Product number: 1340-6069 -
2-port Interface IF 100
Software Winlog.med
USB connection
Colored LEDs signaling programming, readout and incorrect development
Product number: 1340-6064
-
Battery Exchange Set for EBI 11 consisting of: 6 batteries, 3 gaskets with lubricant, screwdriverProduct number: 1100-0120
-
Suitable for 10 battery changes; consisting of 20 batteries, 10 O-rings with grease and changing tool.
Product number: 1100-0125
-
For transport of systainers Fixed seat of the systainers on the trolleyProduct number: 1248-0127
-
- Combined
- Modular construction
- Loggers are safely stored
Product number: 1248-0128 -
Inlay for EBI 11 and interface especially for DAC SetProduct number: 1340-6096-0001
-
ISO calibration 1030-2203 at three standard points for process temperature data loggersProduct number: 1030-2203
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at three points. Calibration service for data logger of the following product series:EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3203
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at four points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3204
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at five points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3205
-
Accredited calibration according to DAkkS guidelines for temperature- validation- and process logger at six points of the following product series: EBI 12, EBI 11, EBI 10, EBI 100.Consisting of an accredited calibration according to DIN EN ISO / IEC 17025.Product number: 1030-3206
- ISO 17665 Sterilization of health care products — Moist heat
- ISO 15883 Washer Disinfectors
- (CE) N° 1935/2004 Materiales y objetos que entran en contacto con los alimentos
- 2014/30/EU The EMC Directive 2014/30/EU regulates the electromagnetic compatibility of equipment