EBI 12-T230 Temperature data logger for Pasteurization
Robust data logger for regular checks in canning factory.
Product information
The data logger is
used for pasteurization control, especially in cans or jars. The sensor needle
is inserted from the outside into the respective Converve and fixed.
Benefits:
- wireless measurement
- Versatile
- large measuring range
- robust data logger
Applications:
- Food Safety
- Pasteurization Control
- PU value determination
- F0 Value calculation
- Sterilisation
Datasheet
Technical Data
- Data memory
- 27.840 measurement values
- Sensor
- Pt 1,000, Class A
- Weight
- ca. 110 g
- Accuracy
- ± 0,5 °C (-55 °C ... -40 °C), ± 0,2 °C (-40 °C … 0 °C), ± 0,1 °C (0 °C … +150 °C)
- Protection class
- IP68
- Dimensions
- (Ø x H) 48 mm x 24 mm - Needle length 50 mm
- Resolution
- 0,01 °C
- Certificate
- Factory Calibration: -20 °C, 0 °C, +60 °C and +134 °C
- Measured variables
- Temperature
- Intervall
- 1 s ... 24 h
- Battery
- Lithium cell, 3.6 V, user replaceable
- Measuring range
- -55 °C … +150 °C
- Storage temperature
- -20 °C ... +70 °C
Accessories
Software & Interfaces
Battery replacement & Maintenance
Carrying Case
Calibration
NEW: Information package "Professional data loggers"
Please fill out the form below and the download will start instantly.
Standards and Directives
- (EC) 1935/2004 Materials and articles which come into contact with foodstuffs
- 2014/53/EU Radio Equipment Directive - RED