Skip to main content Skip to search Skip to main navigation

EBI 12-TP222 Validation logger to measure in autoclaves

Easy to place the bendable metal probe for temperature measurement


Product information

Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its bendable metal probes. The very thin temperature sensors with a diameter of 1.5 mm at the sensor tip enable a very fast response time. With this logger and the AL 126 Endoscope Flow Control Adapter, regular routine checks in RDG-E are easy to perform.

Benefits:

  • bendable metal probe
  • high accuracy
  • Realtime data transmission
  • fast response time
  • sterilizable data logger

Applications:
  • Validation logger for Measurement in the steam sterilizer, WD and endoscope washer

Datasheet

NEW: Information package "Professional data loggers"

You still need a more comprehensive overview of the possibilities and requirements of the data logger, references to the service, certificates or the suitable software? Then read the comprehensive information package for the food sector or the medical/pharmaceutical sector here.
Go to information package

Technical Data

Data memory
2 x 33.440 measurement values
Weight
approx. 120 g
Accuracy
±15 mbar (1 mbar ... 49 mbar) ±10 mbar (50 mbar … 150 mbar) ±15 mbar (151 mbar ... 2.049 mbar) ±10 mbar (2.050 mbar … 2.250 mbar) ±15 mbar (2.251 mbar ... 2.999 mbar) ±10 mbar (3.000 mbar … 3.250 mbar) ±15 mbar (3.251 mbar ... 3.499 mbar) ±20 mbar (3.500 mbar ... 4.000 mbar)
Protection class
IP68
Dimensions
ø 48 mm x 32 mm
Resolution
0,01 °C 1 mbar
Certificate
Factory calibration: 0 °C, +60 °C and +134 °C 100 mbar and 3,100 mbar at +25 °C and 3,100 mbar at +134 °C
Measured variables
Temperature, pressure
Intervall
250 ms ... 24 h
Battery
3.6 V lithium battery exchangeable
Measuring range
Temperature: 0 °C … +150 °C Pressure : 1 … 4.000 mbar
Storage temperature
-20 °C ... +70 °C


NEW: Information package "Professional data loggers"

Please fill out the form below and the download will start instantly.
Information package

Standards and Directives

  • ISO 17665 Sterilization of health care products — Moist heat
  • ISO 15883 Washer Disinfectors
  • 2014/53/EU Radio Equipment Directive - RED
  • (CE) N° 1935/2004 Materiales y objetos que entran en contacto con los alimentos