EBI 12-TP231 Validation logger for steam sterilizers and WD

highliy accurate data logger to measure temperature and pressure


Product information

Validation logger for thermal testing of processes with steam sterilizers and WD's. The logger is characterized by ist high accuracy and ease of use. Of course, this data logger can also be used for routine control in the RUMED.

Benefits:

  • bendable metal probe
  • high accuracy
  • Realtime data transmission
  • fast response time
  • sterilizable data logger

Applications:
  • Validation logger for Measurement in the steam sterilizer, WD and endoscope washer

Datasheet

NEW: Information package "Professional data loggers"

You still need a more comprehensive overview of the possibilities and requirements of the data logger, references to the service, certificates or the suitable software? Then read the comprehensive information package for the food sector or the medical/pharmaceutical sector here.
Go to information package

Technical Data

Data memory
2 x 33.440 measurement values
Sensor
Pt 1,000, class A
Weight
approx. 120 g
Accuracy
±0,1 °C (0 °C ... +120 °C) ±0,05 °C (+120 °C ... +140 °C) ±0,1 °C (+140 °C ... +150 °C)
±15 mbar (1 mbar ... 50 mbar), ±10 mbar (50 mbar … 150 mbar), ±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar), ± 15 mbar (3.250 mbar ... 3.500 mbar),  
± 0,5 % FS (3.500 mbar ... 4.000 mbar)
Protection class
IP68
Dimensions
(Ø x H) 48 mm x 32 mm - Needle length 40 mm
Resolution
Temperature 0,01 °C / pressure 1 mbar
Certificate
Factory Calibration: 0 °C, +60 °C and +134 °C 100 mbar and 3,100 mbar at +25 °C and 3,100 mbar at +134 °C
Measured variables
Temperature, pressure
Intervall
250 ms ... 24 h
Battery
Lithium 3.6 V battery exchangeable
Measuring range
0 °C … +150 °C 1 mbar … 4.000 mbar
Storage temperature
-20 °C ... +70 °C


NEW: Information package "Professional data loggers"

Please fill out the form below and the download will start instantly.
Information package

Standards and Directives

  • ISO 17665 Sterilization of health care products — Moist heat
  • ISO 15883 Washer Disinfectors
  • (EC) 1935/2004 Materials and articles which come into contact with foodstuffs